Behavior of nickel-titanium instruments manufactured with different thermal treatments.

ORCiD

Dr. Ove A. Peters: 0000-0001-5222-8718

Department

Endodontics

Document Type

Article

Publication Title

Journal of Endodontics

ISSN

0099-2399

Volume

41

Issue

1

DOI

10.1016/j.joen.2014.06.005

First Page

67

Last Page

71

Publication Date

1-1-2015

Abstract

INTRODUCTION: The purpose of this study was to investigate if nickel-titanium instruments with similar designs manufactured by different thermal treatments would exhibit significantly different in vitro behavior.

METHODS: Thirty-six instruments each of ProTaper Universal (PTU F1; Dentsply Maillefer, Ballaigues, Switzerland), ProFile Vortex (PV; Dentsply Tulsa Dental Specialties, Tulsa, OK), Vortex Blue (VB, Dentsply Tulsa Dental Specialties), and TYPHOON Infinite Flex NiTi (TYP; Clinician's Choice Dental Products, New Milford, CT) (all size 25/.06) were evaluated. Bending resistance, torsion at failure, and dynamic torsional tests were performed with the instruments (n = 12). Analysis of variance and Tukey post hoc tests were applied.

RESULTS: Flexibility was significantly higher for TYP compared with the other 3 groups (P < .0001). With respect to the maximum torque at failure, PV group showed the highest resistance to twisting (torsional strength) among the analyzed instruments followed by VB, TYP, and PTU. The TYP group exhibited greater angular deflection at failure compared with the other groups (P < .0001). The mean dynamic torque scores during simulated canal preparation were highest for TYP (3.01 ± 0.71 Ncm) and lowest for PV (1.62 ± 0.79 Ncm). However, no significant differences were observed comparing groups PTU with TYP and VB and VB with PV (P > .05). The highest mean forces were recorded with PTU (7.02 ± 2.36 N) and the lowest with TYP (1.22 ± 0.40 N).

CONCLUSIONS: TYP instruments were significantly more flexible than the other instruments tested. The PV group had the highest torsional strength and TYP, despite being the most flexible, showed similar torsional moments to the other instruments, whereas its angular deflection was the highest among the groups.