ORCiD
Dr. Ove A. Peters: 0000-0001-5222-8718
Department
Endodontics
Document Type
Article
Publication Title
Dentomaxillofacial Radiology
ISSN
0250-832X
Volume
41
Issue
8
DOI
10.1259/dmf/81804525
First Page
649
Last Page
655
Publication Date
12-1-2012
Abstract
OBJECTIVES: The various types of cone beam CT (CBCT) differ in several technical characteristics, notably their spatial resolution, which is defined by the acquisition voxel size. However, data are still lacking on the effects of voxel size on the metric accuracy of three-dimensional (3D) reconstructions. This study was designed to assess the effect of isotropic voxel size on the 3D reconstruction accuracy and reproducibility of CBCT data.
METHODS: The study sample comprised 70 teeth (from the Institut d'Anatomie Normale, Strasbourg, France). The teeth were scanned with a KODAK 9500 3D® CBCT (Carestream Health, Inc., Marne-la-Vallée, France), which has two voxel sizes: 200 µm (CBCT 200 µm group) and 300 µm (CBCT 300 µm group). These teeth had also been scanned with the KODAK 9000 3D® CBCT (Carestream Health, Inc.) (CBCT 76 µm group) and the SCANCO Medical micro-CT XtremeCT (SCANCO Medical, Brüttisellen, Switzerland) (micro-CT 41 µm group) considered as references. After semi-automatic segmentation with AMIRA® software (Visualization Sciences Group, Burlington, MA), tooth volumetric measurements were obtained.
RESULTS: The Bland-Altman method showed no difference in tooth volumes despite a slight underestimation for the CBCT 200 µm and 300 µm groups compared with the two reference groups. The underestimation was statistically significant for the volumetric measurements of the CBCT 300 µm group relative to the two reference groups (Passing-Bablok method).
CONCLUSIONS: CBCT is not only a tool that helps in diagnosis and detection but it has the complementary advantage of being a measuring instrument, the accuracy of which appears connected to the size of the voxels. Future applications of such measurements with CBCT are discussed.
Recommended Citation
Maret, D.,
Telmon, N.,
Peters, O. A.,
Lepage, B.,
Treil, J.,
Inglèse, J. M.,
Peyre, A.,
Kahn, J.,
&
Sixou, M.
(2012).
Effect of voxel size on the accuracy of 3D reconstructions with cone beam CT..
Dentomaxillofacial Radiology, 41(8), 649–655.
DOI: 10.1259/dmf/81804525
https://scholarlycommons.pacific.edu/dugoni-facarticles/332
Comments
Republished with permission of British Institute of Radiology, © 2012, from Dentomaxillofacial Radiology.