Analysis of torque and force with differently tapered rotary endodontic instruments in vitro.

ORCiD

Dr. Ove A. Peters: 0000-0001-5222-8718

Department

Endodontics

Document Type

Article

Publication Title

Journal of Endodontics

ISSN

0099-2399

Volume

31

Issue

2

First Page

120

Last Page

123

Publication Date

2-1-2005

Abstract

The aim of this study was to compare torques and forces developed by two sequences of nickel-titanium rotary instruments, using only .04 taper instruments or a combination of .04 and .06 tapers, respectively. Curved canals in plastic blocks were shaped in a torque-testing device; torque, force, contact areas and numbers of working rotations were determined. Analysis of variance and Pearson correlation coefficients were used to compare the two sequences. Mean torque ranged from 0.2 +/- 0.03 to 3.19 +/- 0.28 Ncm and was higher for instrument sizes 25 to 45 in group A. Mean vertical forces ranged from 1.49 +/- 0.54 to 12.0 +/- 3.5 N; there were significant correlations between torque and force as well as between torque and contact areas. Numbers of rotations were lower in the combined .04 and .06 sequence. An instrumentation sequence encompassing various tapers seems to be safer regarding torsional and fatigue failure compared to a sequence that used one single taper. However, the former sequence requires the use of more instruments.